[AWG16a] Towards Kilometric Distance Measurements with Air Refractive Index Compensation
Conférences Internationales sans actes :
JISDM 2016,
vienne,
autriche,
Mots clés: Kilometric distance, optical telemetry,
two-wavelength telemetry, absolute distance meter
Résumé:
The accuracy of kilometric-range distance
measurements by electro-optical instruments
is still limited by the determination of the air
refractive index along the measurement path. Thus,
a sub-millimetric accuracy is not reachable with
classical instruments for such distances. However, a
two-wavelength approach offers a promising route
towards a compensation of air index fluctuations,
and so towards more accuracy. In this paper, the
principle of the two-wavelength absolute distance
meter is explained, an experimental setup working
at 785 nm and 1550 nm is described, and the first
experimental results are presented.